Determination of atenolol in human urine by using HPLC
Özet
A simple high-performance liquid chromatography method has been developed for determination of atenolol in human urine. The method was validated on Ace C-18 column using fluorescence detection. The mobile phase consisted of methanol/water (50:50, v/v) containing 0.1% trifluoroacetic acid. Fluorescence detection was performed at 276 nm (excitation) and 296 nm (emission). The calibration curve was linear between the concentration range of 5-150 ng/mL. Intra- and inter-day precision values for atenolol in human urine were less than 6.09, and accuracy (relative error) was better than 5.50%. The extraction recoveries of atenolol from human urine were between 93.0 and 98.8%. The limits of detection and quantification of atenolol were 1.5 and 5 ng/mL, respectively. Also, the method was successfully applied to a patient with hypertension who had been given an oral tablet of 50 mg atenolol.
Anahtar kelimeler
Atıf için künye
Yılmaz, B., & Arslan, S. (2018). Determination of atenolol in human urine by using HPLC. SEPARATION SCIENCE PLUS, 1(1), 4–10. https://doi.org/10.1002/sscp.201700023
BibTeX
@article{yilmaz2018-22123,
author = {Yılmaz, Bilal and Arslan, Sakir},
title = {Determination of atenolol in human urine by using HPLC},
journal = {SEPARATION SCIENCE PLUS},
year = {2018},
volume = {1},
number = {1},
pages = {4--10},
issn = {2573-1815},
doi = {10.1002/sscp.201700023},
url = {https://akademik.atauni.edu.tr/yayinlar/22123-determination-of-atenolol-in-human-urine-by-using-hplc}
}