Determination of naproxen in human plasma by GC-MS
Özet
This paper describes a GC-MS method for the determination of naproxen in human plasma. Naproxen and internal standard ibuprofen were extracted from plasma using a liquid-liquid extraction method. Derivatization was carried out using N-methyl-N-(trimethylsilyl)trifluoroacetamide. The calibration curve was linear between the concentration range of 0.10-5.0 g/mL. Intra- and interday precision values for naproxen in plasma were <5.14, and accuracy (relative error) was better than 4.67%. The extraction recoveries of naproxen from human plasma were between 93.0 and 98.9%. The LOD and LOQ of naproxen were 0.03 and 0.10 g/mL, respectively. Also, this assay was applied to determine the pharmacokinetic parameters of naproxen in six healthy Turkish volunteers who had been given 220 mg naproxen.
Anahtar kelimeler
Sürdürülebilir Kalkınma Amaçları
-
SKA 3: Sağlıklı Bireyler
Atıf için künye
Yılmaz, B., Sahin, H., & Erdem, A. F. (2014). Determination of naproxen in human plasma by GC-MS. JOURNAL OF SEPARATION SCIENCE, 37(8), 997–1003. https://doi.org/10.1002/jssc.201301105
BibTeX
@article{yilmaz2014-22075,
author = {Yılmaz, Bilal and Sahin, Huseyin and Erdem, Ali Fuat},
title = {Determination of naproxen in human plasma by GC-MS},
journal = {JOURNAL OF SEPARATION SCIENCE},
year = {2014},
volume = {37},
number = {8},
pages = {997--1003},
issn = {1615-9306},
doi = {10.1002/jssc.201301105},
url = {https://akademik.atauni.edu.tr/yayinlar/22075-determination-of-naproxen-in-human-plasma-by-gc-ms}
}