Makale Tam Makale Ulusal 2020

THE BIOMECHANICAL ANALYSIS FOR BIODEGRADABLE PURE MAGNESIUM BONE SCREWS UNDER THREE POINT BENDING AND TORSIONAL TEST

Burak Atik, Yusuf Burak Bozkurt, Yeşim Seçer Kavasoğlu, Yakup Uzun, Halim Kovacı, Ayhan Çelik

Journal of Selcuk-Technic Sayı Özel Sayı · ss. 39–46 · 01.08.2020

Özet

Recently, magnesium and its alloys have been used as the raw material of

degradable implants. In the orthopedic implant group, it is used in the production of

medical tools due to its close modulus of elasticity and mechanical behavior suitable for

bone tissue. In addition, magnesium is completely biodegradable both in biocompatible

and living organisms. The purpose of using a degradable implant within the living

organism is both to perform biomedical functions. It has become even more attractive to

use the biodegradable magnesium screws due to the use of the temporary (nonbiodegradable) implant in the living organism, the need for a secondary surgical operation

to remove it from the body, and the increased risk of complications for the patient.

However, the degradation times of magnesium screws; It is difficult to control the

biological environment, the age and gender of the patients and the implant geometry

Determination of the degradation stage is important for mechanical performance due to

loss of mass and volume in the implant. Furthermore, loss of adhesion performance due

to deterioration of bone screws weakens the mechanical properties of the implant system.Considering this feature of magnesium screws, pure magnesium screws of different origin

were kept in phosphate buffer solution (pH = 7.4) for 1 day, 1,4,12 and 26 weeks. Then,

three-point bending and torsion tests were performed according to ASTM F2502-11

standard to examine the mechanical properties of the screws. In the mechanical tests

applied for both groups, when the three-point bending test and force-displacement curves

were examined, it was observed that the mechanical properties of the 26th week decreased

by about half compared to the 1st day. In the same way, when the torsion-torsion angle

curves were examined in the torsion test, it was observed that the mechanical properties

of the 26th week decreased for both groups.

Sürdürülebilir Kalkınma Amaçları

  • SKA 9: Sanayi, Yenilikçilik ve Altyapı

Atıf için künye

Atik, B., Bozkurt, Y. B., Seçer Kavasoğlu, Y., Uzun, Y., Kovacı, H., & Çelik, A. (2020). THE BIOMECHANICAL ANALYSIS FOR BIODEGRADABLE PURE MAGNESIUM BONE SCREWS UNDER THREE POINT BENDING AND TORSIONAL TEST. Journal of Selcuk-Technic(Özel Sayı), 39–46. http://sutod.selcuk.edu.tr/sutod/article/view/506

BibTeX
@article{atik2020-174336,
  author = {Atik, Burak and Bozkurt, Yusuf Burak and Seçer Kavasoğlu, Yeşim and Uzun, Yakup and Kovacı, Halim and Çelik, Ayhan},
  title = {THE BIOMECHANICAL ANALYSIS FOR BIODEGRADABLE PURE MAGNESIUM BONE SCREWS UNDER THREE POINT BENDING AND TORSIONAL TEST},
  journal = {Journal of Selcuk-Technic},
  year = {2020},
  number = {Özel Sayı},
  pages = {39--46},
  url = {https://akademik.atauni.edu.tr/yayinlar/174336-the-biomechanical-analysis-for-biodegradable-pure-magnesium-bone-screws-under}
}