language English
Article 2022 · 1 atıf

Influence of the sample thickness upon the albedo factors using 59.54 and 662 keV photon energies

Yilmaz, Demet, Akku, Tuba, Kalecik, Sedanur

Özet

The albedo factors (number, energy, and dose) of samples with different thicknesses were measured at a scattering angle of 180 degrees. Ten samples with mass thickness between 0.081 and 0.348 g/cm(2) were prepared from praseodymium powder and twelve samples with the mass thicknesses from 0.068 to 0.620 g/cm(2) were prepared from hafnium powder. The samples were irritated by 59.54 and 662 keV gamma rays emitted from Am-241 and Cs-137 radioactive sources, respectively. The scattered photons were counted by a high purity germanium detector with active area of 200 mm(2). The albedo factors decreased with the atomic number and increased with the mass thickness.

Anahtar Kelimeler
Albedo dose albedo energy albedo number backscattering technique high purity germanium detector BACKSCATTERING
WoS Kategorileri
Chemistry, Analytical Instruments & Instrumentation
Künye
Dergi
INSTRUMENTATION SCIENCE & TECHNOLOGY
Yıl
2022
Cilt
50
Sayı
4
Sayfa
447-455
ISSN
1073-9149
eISSN
1525-6030
DOI
10.1080/10739149.2021.2023175
WoS UID
WOS:000736819500001
Yayını aç (DOI)